A warm part brought straight from machining can change size as it cools. When the change is large relative to the tolerance, temperature history becomes part of the inspection record.

Estimate the effect before testing

For an illustrative 200 mm steel part, assume a linear expansion coefficient of 11.5 µm per metre per °C. A uniform 5 °C difference gives an estimated length change of 11.5 µm. Use the coefficient appropriate to your material and treat the calculation as a first estimate. Gradients through the part and through the machine complicate the result.

Check the part and the machine

Log when the part left production and record temperatures at representative locations. Measure the same feature over time and compare results against an appropriate reference check. Review the machine’s stated environmental limits and the settings for material compensation.

A single room reading gives limited information about a recently machined part. If measurements drift as the part cools, define a release condition supported by the observed stability and the inspection requirements. Repeat the test with typical parts and production conditions before rolling the procedure out.